Implant stability and bone remodeling up to 84 days of implantation with an initial static strain : an in vivo and theoretical investigation

DSpace Repository

Implant stability and bone remodeling up to 84 days of implantation with an initial static strain : an in vivo and theoretical investigation

Show full item record

Files for download

Find Full text There are no files associated with this item..

Facebook

Simple item record

Publication Article, peer reviewed scientific
Title Implant stability and bone remodeling up to 84 days of implantation with an initial static strain : an in vivo and theoretical investigation
Author(s) Anders, Halldin ; Jinno, Yohei ; Galli, Silvia ; Ander, Mats ; Hansson, Stig ; Jacobsson, Magnus ; Jimbo, Ryo
Date 2016
English abstract
Objectives: When implants are inserted, the initial implant stability is dependent on the mechanical stability. To increase the initial stability, it was hypothesized that bone condensation implants will enhance the mechanical stability initially and that the moderately rough surface will further contribute to the secondary stability by enhanced osseointegration. It was further hypothesized that as the healing progresses the difference in removal torque will diminish. In addition, a 3D model was developed to simulate the interfacial shear strength. This was converted to a theoretical removal torque that was compared to the removal torque obtained in vivo. Material and methods: Condensation implants, inducing bone strains of 0.015, were installed into the left tibia of 24 rabbits. Non-condensation implants were installed into the right tibia. All implants had a moderately rough surface. The implants had an implantation time of 7, 28, or 84 days before the removal torque was measured. The interfacial shear strength at different healing time was estimated by the means of finite element method. Results: At 7 days of healing, the condensation implant had an increased removal torque compared to the non-bone-condensation implant. At 28 and 84 days of healing, there was no difference in removal torque. The simulated interfacial shear strength ratios of bone condensation implants at different implantation time were in line with the in vivo data. Conclusions: Moderately rough implants that initially induce bone strain during installation have increased stability during the early healing period. In addition, the finite element method may be used to evaluate differences in interlocking capacity.
DOI http://dx.doi.org/10.1111/clr.12748 (link to publisher's fulltext)
Publisher Wiley
Host/Issue Clinical Oral Implant Research;
ISSN 0905-7161
Pages 1-7
Language eng (iso)
Subject(s) bone condensation
implant stability
in vivo
remodeling
static strain
Medicine
Research Subject Categories::ODONTOLOGY
SCB::Basic Medicine